Designing the Actuator of Hearing Aid Using Spiral Coils and Finite Element Analysis
Article Outline
Abstract
Objective
An electromagnetic vibration transducer was developed to be used as a hearing aid.
Materials and Methods
This type of electromagnetic transducer should have the following characteristics: small in size, high-energy efficiency and suitable frequency bandwidth. In order to find the output of electromagnetic force and to predict the frequency-amplitude characteristics, a finite element middle ear biomechanical model was used to derive the optimal magnetic force of the actuator in this study. First, the electromagnetic transducer was created using a computer-aided design. The air gap between the magnet and coil, input current and vibration force were calculated using finite element analysis simulation.
Results
Simulated results showed that electromagnetic forces under 1.0 mm air gap was about 5–25 dyne depending on the turns of coils and input current of the actuator.
Conclusion
The designed actuator could be used as transducers for middle ear implants or bone anchors in hearing aids.
Keywords: Actuator , Finite element analysis , Hearing aids
No full text is available. To read the body of this article, please view the PDF online.
References
- . Hearing aids: a review of what's new . Otolaryngol Head Neck Surg . 2006;134:1043–1050
- . Earlens tympanic contact transducer: a new method of sound transduction to the human ear . Otolaryngol Head Neck Surg . 1996;114:720–728
- . State of the art on the development of the implantable hearing device for partial hearing loss . Otolaryngol Clin North Am . 1996;29:225–243
- . Current status and critical reflections on implantable hearing aids . Oto Neurotol . 1999;20:409–415
- Three-dimensional reconstruction and modeling of middle ear biomechanics by highresolution computed tomography and finite element analysis . Laryngoscope . 2006;116:711–716
- Biomechanical modeling and design optimization of cartilage myringoplasty using finite element analysis . Audiol Neurootol . 2006;11:380–388
- . Optimal graft thickness for different sizes of tympanic membrane perforation in cartilage myringoplasty: a finite element analysis . Laryngoscope . 2007;117:725–730
- . Measurement of tympanic membrane vibration in 99 human ears . In: Hüttenbrink KB editors. Middle Ear Mechanics in Research and Otosurgery . Dresden, Germany: Dresden University of Technology; 1996;p. 91–93
- Huber A, Ball G, Asai M. The vibration pattern of the tympanic membrane after placement of a total ossicular replacement prosthesis. In: Proceedings of the International Workshop on Middle Ear Mechanics in Research and Otosurgery. Dresden, Germany, 1997:219-22.
- Three-dimensional modeling of middle ear biomechanics and its application . Otol Neurotol . 2002;23:271–280
- . Electromagnetic vibration transducer using polyimide elastic body for implantable middle ear hearing aid . Sensors Actuators A . 2002;97-98:201–207
- . Ein method zur bestimmung der horsch wellenamplituden des trammel fells bei verschieden frequenzen . Skand Arch Physiol . 1935;72:161
- . Ongoing investigations into an implantable electromagnetic hearing aid for moderate to severe sensorineural hearing loss . Otolaryngol Clin North Am . 1995;28:107–120
- Experiences with implantable hearing devices and presentation of a new device . Ann Oto Rhinol Laryngol . 1986;95:60–65
- A contactless electromagnetic implantable middle ear device of the ossicular stimulating type . Ear Nose Throat J . 1994;73:78–90
- Contactless semiimplantable electromagnetic middle ear device for the treatment of sensorineural hearing loss: shortterm and longterm animal experiments . Otolaryngol Clin North Am . 1995;28:121–140
- Electromagnetic implantable middle device of the ossicularstimulating type: principle, designs and experiments . Ann Otol Rhinol Laryngol Suppl . 1988;136:1–16
PII: S1016-3190(08)60022-1
doi:10.1016/S1016-3190(08)60022-1
© 2008 Buddhist Compassion Relief Tzu Chi Foundation. Published by Elsevier Inc. All rights reserved.