Tzu Chi Medical Journal
Volume 20, Issue 2 , Pages 125-129, June 2008

Designing the Actuator of Hearing Aid Using Spiral Coils and Finite Element Analysis

  • Peir-Rong Chen

      Affiliations

    • Department of Otolaryngology, Buddhist Tzu Chi General Hospital, Hualien, Taiwan
    • Department of Medicine, Tzu Chi University, Hualien, Taiwan
  • ,
  • Chia-Fone Lee

      Affiliations

    • Department of Otolaryngology, Buddhist Tzu Chi General Hospital, Hualien, Taiwan
    • Department of Medicine, Tzu Chi University, Hualien, Taiwan
    • Institute of Biomedical Engineering, National Taiwan University, Taipei, Taiwan
    • Corresponding Author InformationCorresponding author. Department of Otolaryngology, Buddhist Tzu Chi General Hospital, 707, Section 3, Chung-Yang Road, Hualien, Taiwan

Received 8 October 2007; received in revised form 24 October 2007; accepted 6 November 2007.

Article Outline

Abstract 

Objective

An electromagnetic vibration transducer was developed to be used as a hearing aid.

Materials and Methods

This type of electromagnetic transducer should have the following characteristics: small in size, high-energy efficiency and suitable frequency bandwidth. In order to find the output of electromagnetic force and to predict the frequency-amplitude characteristics, a finite element middle ear biomechanical model was used to derive the optimal magnetic force of the actuator in this study. First, the electromagnetic transducer was created using a computer-aided design. The air gap between the magnet and coil, input current and vibration force were calculated using finite element analysis simulation.

Results

Simulated results showed that electromagnetic forces under 1.0 mm air gap was about 5–25 dyne depending on the turns of coils and input current of the actuator.

Conclusion

The designed actuator could be used as transducers for middle ear implants or bone anchors in hearing aids.

Keywords:  Actuator , Finite element analysis , Hearing aids

No full text is available. To read the body of this article, please view the PDF online.

 

Back to Article Outline

References 

  1. Kim HH , Barrs DM . Hearing aids: a review of what's new . Otolaryngol Head Neck Surg . 2006;134:1043–1050
  2. Perkins P . Earlens tympanic contact transducer: a new method of sound transduction to the human ear . Otolaryngol Head Neck Surg . 1996;114:720–728
  3. Maniglia AJ . State of the art on the development of the implantable hearing device for partial hearing loss . Otolaryngol Clin North Am . 1996;29:225–243
  4. Hüttenbrink KB . Current status and critical reflections on implantable hearing aids . Oto Neurotol . 1999;20:409–415
  5. Lee CF , Chen PR , Lee WJ , et al.   Three-dimensional reconstruction and modeling of middle ear biomechanics by highresolution computed tomography and finite element analysis . Laryngoscope . 2006;116:711–716
  6. Lee CF , Hsu LP , Chen PR , et al.   Biomechanical modeling and design optimization of cartilage myringoplasty using finite element analysis . Audiol Neurootol . 2006;11:380–388
  7. Lee CF , Chen JH , Chou YF , Hsu LP , Liu TC . Optimal graft thickness for different sizes of tympanic membrane perforation in cartilage myringoplasty: a finite element analysis . Laryngoscope . 2007;117:725–730
  8. Nishihara S , Goode RL . Measurement of tympanic membrane vibration in 99 human ears . In:  Hüttenbrink KB editors. Middle Ear Mechanics in Research and Otosurgery . Dresden, Germany: Dresden University of Technology; 1996;p. 91–93
  9. Huber A, Ball G, Asai M. The vibration pattern of the tympanic membrane after placement of a total ossicular replacement prosthesis. In: Proceedings of the International Workshop on Middle Ear Mechanics in Research and Otosurgery. Dresden, Germany, 1997:219-22.
  10. Gan RZ , Sun Q , Robert KD , et al.   Three-dimensional modeling of middle ear biomechanics and its application . Otol Neurotol . 2002;23:271–280
  11. Park S , Lee KC , Cho JH , Lee SH . Electromagnetic vibration transducer using polyimide elastic body for implantable middle ear hearing aid . Sensors Actuators A . 2002;97-98:201–207
  12. Wilska A . Ein method zur bestimmung der horsch wellenamplituden des trammel fells bei verschieden frequenzen . Skand Arch Physiol . 1935;72:161
  13. Fredrickson J , Coticchia J , Khosla S . Ongoing investigations into an implantable electromagnetic hearing aid for moderate to severe sensorineural hearing loss . Otolaryngol Clin North Am . 1995;28:107–120
  14. Hough J , Vernon J , Johnson J , et al.   Experiences with implantable hearing devices and presentation of a new device . Ann Oto Rhinol Laryngol . 1986;95:60–65
  15. Maniglia A , Ko W , Rosenbaum M , et al.   A contactless electromagnetic implantable middle ear device of the ossicular stimulating type . Ear Nose Throat J . 1994;73:78–90
  16. Maniglia A , Ko W , Rosenbaum  , et al.   Contactless semiimplantable electromagnetic middle ear device for the treatment of sensorineural hearing loss: shortterm and longterm animal experiments . Otolaryngol Clin North Am . 1995;28:121–140
  17. Maniglia A , Ko W , Zhang R , et al.   Electromagnetic implantable middle device of the ossicularstimulating type: principle, designs and experiments . Ann Otol Rhinol Laryngol Suppl . 1988;136:1–16

PII: S1016-3190(08)60022-1